Rapid antigen tests for the diagnosis of a SARS-CoV-2 infection

Welcome! On this webpage we try to summarize the results of manufacturer independent evaluations of antigen detecting rapid diagnostic tests (Ag-RDTs) for SARS-CoV-2.

Background: Since the beginning of the COVID-19 pandemic, numerous Ag-RDTs have appeared on the market. On their homepage, the Federal Institute for Drugs and Medical Devices (BfArM) provides a list of Ag-RDTs [1] which fulfill certain minimum requirements. These requirements were jointly determined by the Paul-Ehrlich-Institute (PEI) and Robert-Koch-Institute (RKI) to include a sensitivity of >70% and a specificity >97% [2].

Manufacturer independent evaluations are important to provide an objective assessment of a test’s accuracy. Thus far only evaluations by the tests’ manufacturers themselves are summarized. Therefore, on this webpage we try to give an overview on the data published by manufacturer independent researchers. Using the linked calculator, it is also possible to convert the shown sensitivity and specificity into the concrete amount of true and false test results.

Manufacturer, TestSensitivity
(CI 95%)
Specificity
(CI 95%)
Ease of useRisk of bias1Study locationIndependence2Source
Abbott, Panbio*85.5%
(78.0-91.2)

91.7%
(89.8-93.4)

79.6%
(67.0-88.8)

73.3%
(62.2-83.8)

75.5%
(69.5-81.5)

86.8%
(79.0-92.0)

72.6%
(64.5-79.9)

81.0%
(69.0–89.8)

71.4%
(63.1-78.7)

86.1%
(78.6-91.7)

90.5%
(87.5–93.6)

66.4%
(59.9-72.2)


60.8%
(46.1–74.2)


48.1%3
(37.4-58.9)


44.6%
(34.3-55.3)
100%
(99.1-100)

98.9%
(97.5-99.6)

100%
(98.7-100)

100%
(not provided)

94.9%
(91.2-98.6)


99.9%
(99.4-100)

100%
(99.7-100)

100%
(99.7-100)

99.8%
(99.4-99.9)

not provided


98.8%
(98.0–99.7)

100%
(99.0 – 100)

not provided


100%
(99.3-100)

100%
(96.3 – 100)
N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A

low concern


intermediate concern

low concern


intermediate concern

low concern


low concern


low concern


low concern


low concern


low concern


intermediate concern

low concern


high concern


low concern


intermediate concern
Switzerland


Spain


Spain


Spain


France


Germany


Netherlands


Aruba


Spain


Switzerland


Spain


Czech Republic


Austria


Spain


Germany

yes


yes


no


yes


yes


yes


yes


yes


yes


yes


no


yes


yes


no


unclear

[10]


[21]


[23]


[24]


[28]


[35]


[33]


[33]


[38]


[39]


[40]


[42]


[45]


[47]


[53]

SD Biosensor / Roche,
Standard Q*
76.6%
(62.8-86.4)

89.0%
(83.7-93.1)

88.7%
(81.3-93.4)

70.6%
(not provided)

81.8%
(71.3-89.6)

92.9%
(86.4-96.9)

84.9%
(79.1-89.4)

70.7%
(not provided)

70%
(not provided)

71.7%
(not provided)

49.4%
(38.9-59.9)


80.3%
(74.9-85.4)
99.3%
(98.-99.6)

99.7%
(98.4-100.0)

97.6%
(95.2-98.8)

100%
(not provided)

99.6%
(97.8-99.9)

not provided


99.6%
(98.6-99.9)

96%
(not provided)

92%
(not provided)

99.5%
(not provided)

100%
(96.3 – 100)

99.1%
(98.6–99.3)
very good


N/A


to come


N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A


N/A

low concern


low concern


low concern


low concern


low concern


low concern


low concern


intermediate concern

high concern


low concern


intermediate concern

intermediate
concern
Germany and England

Switzerland


Brasil


Italy


India


Switzerland


Netherlands


Germany


Uganda


Germany


Germany


Italy

yes


yes


yes


yes


yes


yes


yes


unclear


yes


yes


unclear


unclear

[6], [11]


[10]


[11]


[25]


[30]


[39]


[41]


[43]


[46]


[51]


[53]


[57]

AAZ,
COVID-VIRO
96.7%
(93.5-99.9)

84.1%
(76.9-89.7)
100%
(not provided)

not provided

N/A


N/A

low concern


low concern

France


Switzerland

unclear


yes

[20]


[39]

Bionote, NowCheck89.2%
(81.7-93.9)
97.3%
(94.8-98.6)
N/A

low concern

Brasil

yes

[34]

Coris, Respi-Strip50.0%
(21.5-78.5)


50.0%
(39.5-60.5)


30.2%
(21.7-39.9)


30.0%
(16.7-47.9)
95.9%
(93.5-97.4)


100%
(91.8-100)

100%
(not provided)

100%
(not provided)
insufficient


N/A


N/A


N/A

low concern


intermediate concern

low concern


low concern

Germany and England

France


Belgium


Belgium

yes


unclear


no


yes

[6]


[12]


[26]


[27]

Denka,
QuickNavi
86.7%
(78.6-92.5)
100%
(99.7-100)
N/Alow concernJapanno[60]
Fujirebio,
LUMIPULSE
NP: 81.6%
(71-89.5)

saliva: 41.3%
(30.4-52.8)
NP: 93.8%
(86.2-98)


saliva: 98.6%
(95-99.8)
N/A


N/A

low concern


low concern

Italy


Italy

yes


yes

[59]


[59]

MEDsan,
SARS-Cov-2 Antigen Tests
45.8%
(35.5-56.5)
97.0%
(91.5 – 98.9)
N/A

intermediate concernGermany

unclear

[53]

Nal von Minden, NADAL73.1%
(not provided)
(not provided)N/Ahigh concernGermanyno[58]
Quidel, Sofia SARS Antigen FIA93.8%
(79.9-98.3)

77.0%
(not provided)
100%
(87.5-100)

99.6%
(not provided)
N/A


N/A

high concern


low concern

Chile


United States

yes


unclear

[22]


[43]

RapiGen,
Biocredit
74.4%
(65.8-81.4)

52.0%
(33.5-70.0)


62.0%
(51.0-71.9)


85%
(not provided)

48.1%
(not provided)
98.95%
(97.2-99.6)

100%
(99.7-100)

100%
(88.7-100)

100%
(not provided)

not provided

to come


very good


N/A


N/A


N/A

low concern


low concern


high concern


low concern


high concern

Brasil


Germany


Chile


Nepal


Egypt

yes


yes


no


unclear


yes

[7]


[8]


[14]


[32]


[49]

R-Biopharm,
RIDA QUICK
77.6%
(not provided)

50%
(not provided)
100%
(not provided)

100%
(not provided)
N/A


N/A

low concern


low concern

Germany


Germany

no


no

[54]


[54]

SD Biosensor,
Standard F
69.2%
(53.5-81.4)


77.5%
(69.2-84.1)

90.6%
(75.8-96.8)

47.1%
(37.1-57.1)


62.3%
(55.8-68.4)
96.9%
(95.2-98.0)


97.9%
(95.7-99.0)

100%
(87.5-100)

98.4%
(96.0-99.6)

99.5
(98.0-99.9)
very good


to come


N/A


N/A


N/A

low concern


low concern


high concern


low concern


low concern

Germany


Brasil


Chile


Italy


Czech Republic

yes


yes


yes


yes


yes

[8]


[9]


[22]


[29]


[42]

Shenzen
Bioeasy, FIA
66.7%
(41.7-84.4)


93.9%
(86.5-97.4)

85.0%
(75.6-91.2)
93.1%
(91.0-94.8)


100%
(92.1-100)

100%
(89.0-100)
good


N/A


N/A

low concern


intermediate concern

high concern

Germany and England

Chile


Chile

yes


yes


no

[6]


[13]


[14]

Siemens,
CLINITEST
54.9%
(43.4-65.9)
100%
(96.3 – 100)
N/A

intermediate concernGermany

unclear

[53]

1 The assessment of a possible bias was done using the QUADAS 2 tool. Further information can be found in the methods and sources section.
2 The criteria ‘independence’ evaluates if the study was conducted withoud any incluence by the test’s manufacturer. Further information can be found in the methods and sources section.
3 This study was conducted in a setting where the SARS-CoV-2 postive cases are expected to have a very low viral lode. Thus, the given sensitivity might not be a fair representation of the test’s general accuracy.
* These tests have already been recommended by the WHO.

Follow this link to get to the methods and sources section.

Furthermore, the Paul-Ehrlich-Institute has published a list of Ag-RDTs that they consider to perform to the state of the art. The list can be found following this link.

In addition to the evaluation of Ag-RDTs, we would also like to call attention to further research relevant to this topic.

Correlation of antigen and RNA concentrations:

  • Pollock, N.R., et al., Correlation of SARS-CoV-2 nucleocapsid antigen and RNA concentrations in nasopharyngeal samples from children and adults using an ultrasensitive and quantitative antigen assay. medRxiv. 2020. [36]

Performance of Ag-RDTs in a high-temperature enviroment

  • Haage, V., et al., Impaired performance of SARS-CoV-2 antigen-detecting rapid tests at elevated temperatures. medRxiv. 2021. [61]

Teststrategies:

  • Larremore, D.B., et al., Test sensitivity is secondary to frequency and turnaround time for COVID-19 surveillance. medRxiv. 2020. [15]
  • van Beek, J., et al., From more testing to smart testing: data-guided SARS-CoV-2 testing choices. medRxiv. 2020. [16]
  • Paltiel, A.D., Zheng, A. und Walensky, R.P., Assessment of SARS-CoV-2 screening strategies to permit the safe reopening of college campuses in the United States. JAMA network open 3.7. 2020. [17]
  • Pavelka, M., et al., The effectiveness of population-wide, rapid antigen test based screening in reducing SARS-CoV-2 infection prevalence in Slovakia. medRxiv. 2020. [48]

Alternative sampling-methods:

  • Lindner, A.K., et al., Head-to-head comparison of SARS-CoV-2 antigen-detecting rapid test with self-collected anterior nasal swab versus professional-collected nasopharyngeal swab. medRxiv. 2020. [18]
  • Lindner, A.K., et al., Head-to-head comparison of SARS-CoV-2 antigen-detecting rapid test with professional-collected anterior nasal versus nasopharyngeal swab. medRxiv. 2020. [52]

Evaluation of the Abbott, Panbio in a low prevalence setting for monitoring:

  • Winkel, B.M.F., et al., Screening for SARS-CoV-2 infection in asymptomatic individuals using the Panbio COVID-19 Antigen Rapid Test (Abbott) compared to RT-qPCR. medRxiv. 2020. [50]

Economic modelling:

  • Atkeson, A., et al., Economic Benefits of COVID-19 Screening Tests. medRxiv. 2020. [19]
  • Du, Z., et al., Comparative cost-effectiveness of SARS-CoV-2 testing strategies. Researchgate. 2020. [31]

Get new content delivered directly to your inbox.

%d bloggers like this: