Analytical studies

food plate dawn glass

Besides the clinical studies listed on the homepage, antigen rapid diagnostic tests (Ag-RDTs) are also evaluated in specific laboratory settings. In addition to sensitivity and specificity, such analytical studies analyze the lowest virus concentration an Ag-RDT is able to detect (Limit of Detection, LOD) and whether it incorrectly recognizes any pathogens other than SARS-CoV-2 (cross reactivity). In the following, we would like to give an overview of the results of analytical studies evaluating Ag-RDTs.

Manufacturer,
Test
Study
location
Sensitivity
(CI 95%)
Specificity
(CI 95%)
Cross reactivityLOD [log10
copies/swab]
LOD 50%
(CI 95%) *
LOD 95%
(CI 95%) *
Source
Abbott,
Panbio
Germany


Germany
n/a


n/a
100.00%
(n/a)

n/a
– RSV-A


n/a
6.551


6.07
5.61
(5.27-5.95)

n/a
7.45
(6.79-8.20)

n/a
[37]


[61]
Abbott,
BinaxNOW
Germanyn/an/an/a4.6-4.92n/an/a[56]
Biocredit,
Rapigen
Germanyn/a100.00%
(n/a)
none10.219.51
(8.84-12.26)
11.10
(9.71-17.01)
[37]
BIONOTE,
NowCheck
Germanyn/an/an/a6.0n/an/a[61]
Boditech Med,
ichroma
Germanyn/an/an/a5.91n/an/a[61]
Coris,
Respi-Strip
Germany


Germany
n/a


n/a
100.00%
(n/a)

n/a
none


n/a
7.461


7.90
7.60
(7.37-7.82)

n/a
8.36
(8.00-8.76)

n/a
[37]


[61]
Edinburgh Genetics,
ActivXpress
Germanyn/an/an/a6.15n/an/a[61]
Genedia,
COVID-19 Ag
Germanyn/an/an/a7.66n/an/a[61]
Healgen,
Rapid Test
Germanyn/a91.42%
(n/a)
– Adenovirus
– Entero / Rhinovirus
-Influenzavirus A H1
– Influenzavirus A H3
– Parainfluenzavirus 1
– Parainfluenzavirus 2
– Parainfluenzavirus 3
6.3714.48
(3.41-5.32)
7.27
(6.27-8.40)
[37]
JOYSBIO (Tianjin)
COVID-19 Ag-RTK
Germanyn/an/an/a6.15n/an/a[61]
R-Ratiopharm,
RIDA Quick
Germanyn/a82.86%
(n/a)
– Influenzavirus A H1
– Influenzavirus A H3
6.3215.40
(4.99-5.77)
7.22
(6.57-7.96)
[37]
NAL von minden,
NADAL
Germanyn/a97.12%
(n/a)
none6.9717.19
(6.97-7.43)
7.87
(7.52-8.23)
[37]
SD Biosensor /
Roche,
Standard Q
Germany


Germany
n/a


n/a
97.12%
(n/a)

n/a
– Parainfluenzavirus 3


n/a
6.781


6.0
5.64
(5.23-6.00)

n/a
7.68
(6.96-8.50)

n/a
[37]


[61]
Shenzen Bioeasy,
FIA
Germanyn/an/an/a6.85n/an/a[61]
Siemens Healthineers,
Clinitest
Germanyn/an/an/a6.04n/an/a[61]

* The values “LOD 50% and 95% by viral load” were only determined in study [37]. However, their sample preparation included a step of dilution that was not taken into account for the “LOD 50% and 95% by viral load” values. Thus, these values are eight-fold lower. Still, they can be usefull to estimate the confidence interval and as a rough guide on how far the viral load can decrease to still be detected in 50% of the samples. All values are shown in log10 SARS-CoV-2 RNA copies per swab.

1 At the given LOD, the authors accepted a 5% false-negative rate. Thus, the true LOD could vary slightly.

2 The authors allowed no false-negatives. Only a LOD-range could be provided as the estimation of virus concentration of the test specimen was done using two different PCR methods. The true value lies in between. 