Besides the clinical studies listed on the homepage, antigen rapid diagnostic tests (Ag-RDTs) are also evaluated in specific laboratory settings. In addition to sensitivity and specificity, such analytical studies analyze the lowest virus concentration an Ag-RDT is able to detect (Limit of Detection, LOD) and whether it incorrectly recognizes any pathogens other than SARS-CoV-2 (cross reactivity). In the following, we would like to give an overview of the results of analytical studies evaluating Ag-RDTs.
| Manufacturer, Test | Study location | Sensitivity (CI 95%) | Specificity (CI 95%) | Cross reactivity | LOD [log10 copies/swab] | LOD 50% (CI 95%) * | LOD 95% (CI 95%) * | Source |
|---|---|---|---|---|---|---|---|---|
| Abbott, Panbio | Germany Germany | n/a n/a | 100.00% (n/a) n/a | – RSV-A n/a | 6.551 6.07 | 5.61 (5.27-5.95) n/a | 7.45 (6.79-8.20) n/a | [37] [61] |
| Abbott, BinaxNOW | Germany | n/a | n/a | n/a | 4.6-4.92 | n/a | n/a | [56] |
| Biocredit, Rapigen | Germany | n/a | 100.00% (n/a) | none | 10.21 | 9.51 (8.84-12.26) | 11.10 (9.71-17.01) | [37] |
| BIONOTE, NowCheck | Germany | n/a | n/a | n/a | 6.0 | n/a | n/a | [61] |
| Boditech Med, ichroma | Germany | n/a | n/a | n/a | 5.91 | n/a | n/a | [61] |
| Coris, Respi-Strip | Germany Germany | n/a n/a | 100.00% (n/a) n/a | none n/a | 7.461 7.90 | 7.60 (7.37-7.82) n/a | 8.36 (8.00-8.76) n/a | [37] [61] |
| Edinburgh Genetics, ActivXpress | Germany | n/a | n/a | n/a | 6.15 | n/a | n/a | [61] |
| Genedia, COVID-19 Ag | Germany | n/a | n/a | n/a | 7.66 | n/a | n/a | [61] |
| Healgen, Rapid Test | Germany | n/a | 91.42% (n/a) | – Adenovirus – Entero / Rhinovirus -Influenzavirus A H1 – Influenzavirus A H3 – Parainfluenzavirus 1 – Parainfluenzavirus 2 – Parainfluenzavirus 3 | 6.371 | 4.48 (3.41-5.32) | 7.27 (6.27-8.40) | [37] |
| JOYSBIO (Tianjin) COVID-19 Ag-RTK | Germany | n/a | n/a | n/a | 6.15 | n/a | n/a | [61] |
| R-Ratiopharm, RIDA Quick | Germany | n/a | 82.86% (n/a) | – Influenzavirus A H1 – Influenzavirus A H3 | 6.321 | 5.40 (4.99-5.77) | 7.22 (6.57-7.96) | [37] |
| NAL von minden, NADAL | Germany | n/a | 97.12% (n/a) | none | 6.971 | 7.19 (6.97-7.43) | 7.87 (7.52-8.23) | [37] |
| SD Biosensor / Roche, Standard Q | Germany Germany | n/a n/a | 97.12% (n/a) n/a | – Parainfluenzavirus 3 n/a | 6.781 6.0 | 5.64 (5.23-6.00) n/a | 7.68 (6.96-8.50) n/a | [37] [61] |
| Shenzen Bioeasy, FIA | Germany | n/a | n/a | n/a | 6.85 | n/a | n/a | [61] |
| Siemens Healthineers, Clinitest | Germany | n/a | n/a | n/a | 6.04 | n/a | n/a | [61] |
* The values “LOD 50% and 95% by viral load” were only determined in study [37]. However, their sample preparation included a step of dilution that was not taken into account for the “LOD 50% and 95% by viral load” values. Thus, these values are eight-fold lower. Still, they can be usefull to estimate the confidence interval and as a rough guide on how far the viral load can decrease to still be detected in 50% of the samples. All values are shown in log10 SARS-CoV-2 RNA copies per swab.
1 At the given LOD, the authors accepted a 5% false-negative rate. Thus, the true LOD could vary slightly.
2 The authors allowed no false-negatives. Only a LOD-range could be provided as the estimation of virus concentration of the test specimen was done using two different PCR methods. The true value lies in between.